Precautions during the use of trace gas analyzer

    1. The airtightness of the sampling line of the analyzer

    The supporting pipeline of the trace oxygen analyzer should be sealed. A very small leak will cause the oxygen in the ambient air to diffuse in, resulting in a high measurement value.

    Although in the measurement, the sample gas pressure is greater than the ambient pressure, but the oxygen in the sample gas is a micro-scale. According to Ferrari's law, the partial pressure of oxygen is proportional to its volume content. The atmosphere contains about 21% oxygen, and PPM. The oxygen partial pressure of the sample gas of the calculated concentration differs by about 10,000 times, so the partial pressure of the trace oxygen in the gas sample is much lower than the partial pressure of oxygen in the atmosphere. When the leak occurs, the oxygen in the atmosphere will rapidly spread from the leaking part. Come in.

    Secondly, the sampling pipeline should be as short as possible, the joints should be as small as possible, and the joints and valves should be sealed well. After the pipeline is connected, the air tightness inspection should be done.

    Request for air tightness test: Under 0.25 MPAm test pressure, the pressure drop is not more than 0.01 MPA for 30 minutes.

    2, the selection of sampling material for the trace oxygen analyzer

    The pipeline material is basically made of copper or stainless steel pipeline, and the secondary PTFE tube is selected. Forbidden selection of pipes such as latex tubes and white rubber tubes, the air tightness and material resistance to penetration are too poor, and the measurement of trace oxygen is too much error under the standard measurement pressure. The outer diameter of the pipeline is usually 6mm or 1/4IN. We also choose 3mm or 1/8IN. In short, choose stainless steel tube, clean and degrease, and keep the inner wall of the tube clean. For the analysis of trace level (<1PPMV) oxygen, The inner wall polished stainless steel tube should be selected. The selected valve and joint should have a dead volume as small as possible.

    3, the time of sampling and analysis of the elements of temperature

    In order to prevent the moisture in the sample from condensing and condensing on the tube wall, forming a dissolution and absorption of trace oxygen, the sampling pipeline should be insulated or heat-insulated according to the condition. When detecting trace oxygen in liquid nitrogen, special attention should be paid to heating measures. Otherwise, since the boiling point of oxygen is lower than the boiling point of nitrogen by 13 degrees, the sample gas is not vaporized evenly, which will make the measured value seriously lower.

    4. The closer the sampling point is to the analyzer, the better.

    The measurement position of the trace oxygen analyzer should be as close as possible to the measurement point, so as to avoid excessive pipelines and excessive unreliable factors, which affect the reliability of the measurement data.

    5, the sample gas sampling process should be clean and clean

    The sample gas should not contain oil components or solid particles to avoid clogging and contamination of the membrane.

    6, the sample gas can not contain corrosive gases

    The sample gas should not contain sulfide, phosphide or acid gas components. These components pose a hazard to fuel cells, particularly alkaline fuel cells.

    Lithium Base Grease

    Lithium Base Grease,Molybdenum Disulfide Grease,Complex Lithium Base Grease,Extreme Pressure Lithium Grease

    PUYANG XINYE SPECIAL LUBRICATING OIL AND GREASE CO.,LTD , https://www.xinyelubricant.com

    Previous Post: The following points should be achieved when drilling rigs in normal drilling of pile foundations
    Next Post: SolidWorks solid part modeling, completed in 10 steps
    Home
    Recent Posts
    • New Design Cramic Clutch Button
    • Auto Clutch Button with High Quality (XSCF001)
    • HOWO 336HP 6X4 20m3 25ton Heavy Dumper Tipper Tr…
    • Suzuki YAMAHA Bajaj Moyorcycle Spare Parts for H…
    • Pesticides: production and sales activity will c…
    • Incorporation of Changan Group Changhe Suzuki wi…
    • New transgenic wheat can resist fungi
    • 130 Zhengyue China Factory Plastic Injection Mou…
    • 12.5X1.75X2.25 20X1.75 BMX Bicycle Tire
    • Chock Type a
    • Clutch Disc for Benz 601, 602 Bus OEM: 186151533…
    • Clutch Facing with Akebono Appearance (GREEN TYP…
    • High Quality Ceramic Clutch Button
    • Clutch Button for Auto Parts (XSCF005F)
    • Clutch Button for Mack Truck, Automobile Clutch …
    • CNC Machining Aluminum Cap Non Standard Screw Tu…
    • Steel Auto Parts for All Models and All Spare Pa…
    • Industrial naphthalene is expected to return to …
    • Subsidy for new energy vehicles drastically retr…
    • South Korea NK wins $10 million core parts order